Directions for Use -- Infant/Pediatric Patient Circuits

The following steps describe how to attach a Pediatric/Adult patient circuit to a ZOLL
Ventilator and verify that it is working correctly.
Inspiratory Line
Pressure Line
(Green)
Exhalation
Valve Line
(Clear)
(Clear)
Figure D-6 Infant/Pediatric/ Patient Circuit
1. Connect inspiratory line to the ventilator's gas outlet.
Connect the opposite end of the green pressure line to the airway pressure transducer fitting
2.
on the ventilator. Make sure that the connection is secure by fully pushing the connector to
the metal panel.
Connect the opposite end of the clear exhalation valve line to the exhalation valve fitting on
3.
the ventilator.
Verify that all tubing and hose connections are securely tight.
4.
Cycle the ventilator in accordance with operating instructions with the ventilator connected
5.
to a test lung prior to patient use; check for leaks and occlusions. Ensure that gas properly
flows through the exhalation valve.
Note:
D-10
Exhalation
Valve
Cut off the cuff connector on the green pressure line as shown in the following figure,
and connect securely to the transducer fitting:
Cut Here
ZOLL Medical Corporation recommends that, on a daily basis, you examine the
patient circuit for damage or wear, such as cracking, discoloration, or disfigurement. If
there is any sign of physical degradation or if the unit is indicating breathing problems,
replace the patient circuit
www.zoll.com
Note: Wipe a pair of
scissors with isopropyl
alcohol or equivalent
disinfectant and allow
to dry before cutting.
906-0731-01-05 Rev. C
Patient
Connection
Elbow with
Pressure Tap