A
LARMS
1173
Self Check Failure
Alarm occurs when communication fails between one of the subcomponents and the host processor. If this
failure occurs, the user should manually ventilate the patient, replace the ventilator and contact the service
center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, Backup Ventilator Started, **Contact Service
1174
Self Check Failure
Alarm occurs when the device is not able to calibrate the one or more transducers and is no longer able to
operate safely. If this failure occurs, the user should manually ventilate the patient, replace the ventilator and
contact the service center for additional information.
Mitigation/Info: Pneumatic Sensor: Transducer, Manually Ventilate Patient, Restart Ventilator, **Contact Ser-
vice Center**
1175
Self Check Failure
Alarm triggers when the internal communication bus and the host are not able to communicate with the sub-
assemblies. If this failure occurs, the user should manually ventilate the patient, replace the ventilator and
contact the service center for additional information.
Mitigation/Infor: Internal COMM, Manually Ventilate Patient, **Contact Service Center**
1176
Self Check Failure
Alarm triggers when the calibration file fails its integrity check. The user should manually ventilate the patient,
replace the ventilator and contact the service center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, **Contact Service Center**
1420
Self Check: Complete Power Failure
Alarm triggers when power is lost from both the internal battery and an external source during operation.
When this occurs, the LCD blanks (no power for operation), the audible alarm pulses rapidly, and the visual
alarm flashes rapidly. This alarm will last approximately two minutes. If the device can be recharged after the
failure and there are no other issues it can be returned to service. If there are any questions, contact the ser-
vice center for additional information.
1430
Drained Battery
Alarm triggers when the internal battery power drops below the amount required to provide ventilation and
external power is not connected. When this occurs there is enough power to operate the user interface and
provide information to the user. The user should be manually ventilate the patient while an external source of
power is sought. To cancel the alarm and begin operation with external power the device must be turned off
and then back on.
Mitigation/Info: Manually Ventilate Patient, Connect External Power, **Contact Service Center**
1471
Self Check Failure
Alarm triggers when the device is no longer able to communicate with the User Interface Module (UIM) and
the interface controls. When this occurs ventilation continues at the current settings or the backup mode set-
tings and the high priority alarm sounds. The user should manually ventilate the patient, replace the ventilator
and contact the service center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, **Contact Service Center**
1472
Self Check Failure
Alarm triggers when the device is no longer able to communicate with the Smart Pneumatic Module (SPM).
When this occurs ventilation continues at the current settings or the backup mode settings and the high prior-
ity alarm sounds. The user should manually ventilate the patient, replace the ventilator and contact the ser-
vice center for additional information.
Mitigation/Info: Internal COMM, Manually Ventilate Patient, **Contact Service Center**