Fio2 (Fraction Of Inspired Oxygen) -- Oxygen Delivery Management - ZOLL AEV Operator's Manual

Hide thumbs Also See for AEV:
Table of Contents

Advertisement

U
ZOLL V
SING THE
ENTILATOR
FIO2 (Fraction of Inspired Oxygen) -- Oxygen Delivery
Management
Pressing the FIO
it. There are no adjustable secondary parameters. The default values at start up is 21% whether
oxygen is present or not. If an FIO
settings, the unit start ups with that saved FIO
high-pressure oxygen is not present, the unit starts up with FIO
PRESSURE LOW alarm is not activated. The secondary display in the parameter window is O
1
Use
. This is the flow (liters/min) of high pressure oxygen used by the unit to support the
patient at the current settings. O
sign next to the FIO
oxygen use in the O
The following table lists the options and ranges for the FIO2 parameter:
Parameter Window
Primary Value
Secondary Values
Alarm Limits
Measured Values
FiO2 Context Menu
O2 Reservoir
1. O
Use = ((FIO
2
volume is the actual minute volume (controlled and spontaneous breaths * tidal volume).
4-12
parameter button highlights the current FIO
2
2
Reservoir mode is indicated on the display with a plus "+"
2
value when this mode is active. (The "O
2
Reservoir.)
2
Options /
Range
FiO2
21 to 100
%
Not Applicable
Not Applicable
0 to 99.9
O2 Use (L/min)
Off / On
Default : off
-0.21)/0.79)*Minute Volume where FIO
2
www.zoll.com
value greater than 21% is saved and used for Power Up
value if high-pressure oxygen is present. If
2
Availability/Notes
All breaths are delivered from 
the compressor at 21%
All breaths are delivered from 
the High Pressure O2 Source at 
100%
Shows when High Pressure
Oxygen Supply is present.
"+" icon indicates when "on" for 
low flow oxygen.
is represented as a fraction and minute
2
value and enables you to adjust
2
= 21% and O
SUPPLY
2
2
Use" value does not include
2
906-0731-01-05 Rev. C
2
Models
All
All
All
All
All