Fresh Gas Intake and Attachments

The fresh gas intake, which is located on the side of the ventilator, allows ambient air into the
unit's internal compressor. The intake also acts as an anti-asphyxia value that enables the
patient to breath ambient air should the ventilator fail.
The fresh gas intake contains a particulate filter and permits the operator to connect either a
bacteria/viral or a chemical/biological filter depending on ambient conditions
ZOLL provides an O2 Reservoir Kit to allow for low flow oxygen supply to the ventilator. An
oxygen concentrator source provides oxygen to the O2 reservoir.

Top Panel

The ZOLL Ventilator's top panel appears as follows:
Transducer
(Patient Airway
Pressure)
Exhalation
Valve
The oxygen hose, ventilator circuit, external power, and pulse oximeter attach to the top panel
of the ventilator. The USB port is only used when servicing the unit.
2-10
High Pressure
Oxygen Input
Gas Output
www.zoll.com
External Power Input
Pulse Oximeter Connector
+ USB Connector
906-0731-01-05 Rev. C
Fresh Gas/
Emergency
Air Intake