HR (Heart Rate)

The HR (Heart Rate) parameter window displays the patient's heart rate when the pulse
oximeter is working and the sensor is attached.
Pressing the HR parameter button highlights the High Heart Rate alarm limit and enables its
value to be changed. Pressing the HR button a second time highlights the current value of the
Low Heart Rate Alarm limit and enables its value to be changed. Both limits are adjustable by
1 b/min. The default value at start up for the high alarm limit is 120 BPM (Beats Per Minute);
the low alarm limit is 40 BPM.
The following table gives the options and ranges for the HR parameter:
Parameter Window
Primary Value
Secondary
Values
Alarm Limits
Measured
Values
HR Context Menu (note same as SpO2 Context Menu)
4-16
 
Options /
Range
HR
0 to 240
%
Not Applicable
High Limit
80 to 240,
Off
Low Limit
30 to 79,
Off
Pleth Waveform
www.zoll.com
Availability/Notes
Measurement - Heart Icon blinks
at the beat rate.
Models
All
906-0731-01-05 Rev. C