d.
e.
f.
906-0731-01-05 Rev. C
Alarm Icons: For each active alarm, an alarm bell appears. When multiple alarms
are active, the number of bells corresponds to the number of alarms. The alarm in
the AMC is displayed as the solid bell. To view each active alarm, turn the Dial to
scroll through all active alarms. The plot icon is also in this list. It allows you to see
the current waveform to better assess the nature of the failure. A maximum of six
alarms can be displayed without the plot icon.
Service Code: Each alarm has a 4 digit number associated with it, which helps the
operator indicate the specific alarm when communicating with technical support.
The service codes appear in the following format:
1###
High Priority Alarm
2###
Medium Priority Alarm
3###
High Priority Alarm
Attention Warning Icon: Identifies the severity of the alarm: Low, Medium, or
High priority. See the Symbols table in Chapter 1 for the appearance of the warning
triangle for each of these three alarms.
ZOLL Ventilator Operator's Guide
5-3