Connection
Expiratory Line (Clear)
Oxygen In
Exhaust
Specifications
Pediatric/Adult, 6 ft Patient Circuit
The Pediatric/Adult, 6 ft patient circuit has the following specifications:
Internal Diameter: 22 mm
•
Inspiratory Resistance: R
•
Expiratory Resistance: R
•
Tubing Compliance: C
•
Dead Space: 22 ml
•
Maximum Working Pressure: 100 cm H
•
Pediatric/Adult, 12 ft Patient Circuit
The Pediatric/Adult, 12 ft patient circuit has the following specifications:
Internal Diameter: 22 mm
•
Inspiratory Resistance: R
•
Expiratory Resistance: R
•
Tubing Compliance: C
•
Dead Space: 22 ml
•
Maximum Working Pressure: 100 hPa (cm H
•
Infant/Pediatric, 6 ft Patient Circuit
The Infant/Pediatric, 6 ft patient circuit has the following specifications:
Internal Diameter: 10 mm
•
Inspiratory Resistance: R
•
Expiratory Resistance: R
•
Tubing Compliance: C
•
Dead Space: 4.2 ml
•
Maximum Working Pressure: 100 hPa (cm H
•
9650-002360-01 Rev. A
Symbol on the Ventilator
@ 30 Lpm: 0.01 hPa/l/min
INSP
@ 30 Lpm: 0.10 hPa/l/min
EXP
@ 60 hPa: 1.6 ml/cm H
T
0 (hPa)
2
@ 30 Lpm: 0.02 hPa/l/min
INSP
@ 30 Lpm: 0.10 hPa/l/min
EXP
@ 60 hPa: 2.8 ml/hPa
T
@ 15 Lpm: 0.11 hPa/l/min
INSP
@ 15 Lpm: 0.17 hPa/l/min
EXP
@ 60 hPa: 0.5 ml/hPa
T
ZOLL Z Vent Ventilator Operator's Guide
Description
Exhalation Valve
High Pressure Oxygen
Outlet
Do Not Occlude
0 ml/(hPa)
2
O)
2
O)
2
Product Overview
2-15