Features And Accessories - Tektronix P5200 Instruction Manual

High voltage differential probe
Hide thumbs Also See for P5200:
Table of Contents

Advertisement

Features and Accessories

The P5200 High Voltage Differential Probe provides a safe means of
measuring circuits with floating potentials up to 1,000 V
earth ground and up to 1,300 V (DC + peak AC) differential. This probe
must be connected to an oscilloscope or other measurement instrument
in which the input BNC connector is at ground potential. The probe is a
safe alternative to the extremely dangerous practice of disconnecting
the oscilloscope ground to achieve a floating measurement.
The P5200 probe allows clear and accurate measurements of high-speed
transitions and provides excellent common-mode rejection of noisy
signals. Both inputs have high impedance and low capacitance. Because
of these features, the probe can safely measure the fast voltage transients
in switching power devices such as IGBTs, power MOSFETs, thyristors,
GTOs, and bipolar transistors without damaging these devices.
Other applications for the P5200 probe include testing high-voltage
motor control circuits and line connected circuits in switch-mode power
supplies.
WARNING.
Voltage Differential Probe with oscilloscopes that have floating inputs
(isolated inputs), such as the Tektronix TPS2000 series oscilloscopes and
THS700 series oscilloscopes. The P5200 High Voltage Differential Probe
requires an oscilloscope or other measurement instrument with a grounded
input.
WARNING.
leads are connected to circuits above the voltage and frequency limits
specified. (See page 9, Safety limits.) Use only probe accessories that are
rated for the application.
CAUTION.
1,000 V
RMS
AC) between the leads.
P5200 Instruction Manual
To avoid an electric shock, do not use the P5200 High
To avoid RF burns, do not handle the probe while the input
Do not use the P5200 High Voltage Differential Probe above
CAT II from ground on either input or ± 1,300 V (DC + peak
CAT II from
RMS
1