Typical Characteristics - Tektronix P5200 Instruction Manual

High voltage differential probe
Hide thumbs Also See for P5200:
Table of Contents

Advertisement

Specifications
Table 2: Warranted Electrical Characteristics (cont.)
Characteristics
Maximum Nondestructive
Input Voltage
Range Accuracy
Temperature
Humidity

Typical Characteristics

The typical characteristics in the following tables describe typical, but
not guaranteed, performance.
Table 3: Typical Electrical Characteristics
Characteristics
Rise Time
Bandwidth
AC Common-Mode Rejection
Ratio (20-30 °C, <70% RH)
Noise (measured tangentially)
Input Impedance and
Capacitance
DC Output Drift
Propagation Delay
Table 4: Typical Mechanical Characteristics
Characteristics
Dimensions, Case
Dimensions, Input Leads
16
Description
1/500 and 1/50 differential: ±1.3 kV (DC + peak AC)
1/500 and 1/50 common mode: ±1 kV
±3% at 20-30 °C, <70% RH after 20 minute warm up
Operating: 0 to 40 °C
Nonoperating: –30 to +70 °C
Operating: 25 to 85% RH, +25 to +35 °C
Nonoperating: 25 to 85% RH, +25 to +60 °C
Description
<14 ns in 1/500 range setting
DC to 25 MHz (–3dB) in 1/500 range setting
60 Hz: >10,000:1, –80 dB
100 kHz: >300:1, –50 dB
1 MHz: >300:1, –50 dB
<4 mV
RMS
8 MΩ, 3.5 pF between inputs
4 MΩ, 7 pF between each input and ground
±0.5 mV/°C
20 nS
Description
185 mm × 66 mm × 32 mm
(7.2 in × 2.6 in × 1.3 in)
46 cm (18 in)
CAT II
RMS
P5200 Instruction Manual

Advertisement

Table of Contents
loading

Table of Contents