Tektronix P5200 Instruction Manual page 24

High voltage differential probe
Hide thumbs Also See for P5200:
Table of Contents

Advertisement

Operating Basics
Minimizing Risk of RF Burn (probe leads)
WARNING.
the leads are connected to a source that is above the voltage and frequency
limits. (See page 9, Safety limits.) The area above these limits poses a
risk of radio frequency (RF) burns.
If you need to use the probe within the risk area for RF burn, power off
the source before connecting or disconnecting the probe leads.
Maximum Input Limits
CAUTION.
not apply a voltage that is more than 1,000 V
input and ground or more than 1,300 V (DC + peak AC) between the two
inputs. Above 3 MHz, the voltage limit decreases as frequency increases.
(See page 9, Safety limits.) The input limit applies to both the 50X and
500X settings.
8
To avoid personal injury, do not handle the probe leads when
To avoid damaging the input circuitry of the P5200 probe, do
CAT II between either
RMS
P5200 Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P5200 and is the answer not in the manual?

Questions and answers

Table of Contents