Figure 7: Slide probe leads onto the binding posts
WARNING.
disabled before setting the volage above 30Vrµs.
11. Set the probe range button on the probe to
generator for a 50 V and 1 kHz standard amplitude (AUX, Square
wave, 1 MΩ load) output.
WARNING.
do not touch exposed metal parts after the generator output is enabled.
12. Enable the output of the generator.
13. Record the DC amplitude of the square wave. Divide into the
amplitude of just the oscilloscope (refer to step 6). Verify that only
the probe gain accuracy is ±3%.
14. Set the calibration generator for 5 V standard amplitude (square
wave) output. Set the probe range button on the probe to
15. Record the DC amplitude of the square wave. Divide into the
amplitude of the oscilloscope (refer to step 6). Verify that only
the probe gain accuracy is ±3%.
16. Reduce the amplitude on the generator to minimum, then disable
the generator output and disconnect the setup.
P5200 Instruction Manual
To reduce the risk of shock, ensure the generator output is
Generator produces hazardous voltage. To avoid risk of shock,
Performance Verification
/
(out). Set the
1
500
/
1
50
(in).
25