Remove The Main Board - Welch Allyn 53000 Service Manual

Vital signs monitor 300 series
Hide thumbs Also See for 53000:
Table of Contents

Advertisement

64
Disassembly procedure

Remove the main board

1. Disconnect the temperature module flex cable from J12 on the main board.
Temperature module flex cable
Main board J12 connector
2. Remove the four screws securing the main board to the rear chassis.
Welch Allyn VSM 300 Series
Unlatch the ZIF connector before
removing or inserting the flex
cable