Functional Verification Procedure; Reference Illustration; System/Power; Setup - Welch Allyn 53000 Service Manual

Vital signs monitor 300 series
Hide thumbs Also See for 53000:
Table of Contents

Advertisement

Service manual

Functional verification procedure

Reference illustration

Systolic
window
Diastolic
window
Adult
Pediatric
Neonatal
Message
window

System/power

Note

Setup

If the monitor is configured with the temperature option, connect the temperature
probe and insert it into the probe well.
If you are using an optional Welch Allyn Model Calibration Tester, plug it in and set it to
the values that match your tester model:
Model
9600
9600 Plus
If the monitor is configured with the SpO
Battery charge and beeper
1. With the monitor turned off, disconnect the power adapter from the monitor.
2. Verify that the charge LED
3. Connect the power adapter. The monitor emits a single beep tone.
4. Verify that the charge LED
SYS
DIA
Power
Review
On/Off
patient
data
Other than the optional NIBP overpressure test
here must be performed as part of a complete functional verification procedure.
Temperature
96.4 °F
(35.8 °C)
96.8 °F
(36.0 °C)
/min
SpO
%
2
Set
Start/
Set
NIBP
stop
alarm
interval
NIBP
limits
(page
option, connect the SpO
2
is off.
is on.
Functional verification
Fahrenheit
Celsius
Print patient data
Change settings
Monitored
temperature
Increase/Decrease
Forward/Back
Battery charged
Battery charging
Battery low
Silence
alarms
18), the tests described
sensor.
2
11