Functional verification
6. Verify that the reading displayed on the monitor is as follows:
Model
9600
9600 Plus
7. Remove the probe from the tester and replace it in the probe well.
8. Disconnect the tester.
Temperature
106.0 °F ±0.4 °F
(41.1 °C ±0.2 °C)
105.8 °F ±0.4 °F
(41.0 °C ±0.2 °C)
Welch Allyn VSM 300 Series