Valve And Pump Current, Inflation, And Deflation Tests - Welch Allyn 53000 Service Manual

Vital signs monitor 300 series
Hide thumbs Also See for 53000:
Table of Contents

Advertisement

Service manual
4. Press
approximately 80 mmHg (10.7 kPa), and then settles at a slightly lower pressure level.
Wait a few seconds for the pressure to stabilize.
5. Verify that the value displayed in SYS is within 3 mmHg (0.4 kPa) of the value
displayed on the digital pressure meter.
6. Press
approximately 150 mmHg (20 kPa), and then settles at a slightly lower pressure level.
Wait a few seconds for the pressure to stabilize.
7. Verify that the value displayed in SYS is within 3 mmHg (0.4 kPa) of the value on the
digital pressure meter.
8. Press
mmHg (40 kPa), and then settles at a slightly lower pressure level. Wait a few
seconds for the pressure to stabilize.
9. Verify that the value displayed in SYS is within 6 mmHg (0.8 kPa) of the value on the
digital pressure meter.

Valve and pump current, inflation, and deflation tests

Note
To test pump current
1. Put the monitor in Service mode:
a. Power the monitor off.
b. Power the monitor on while pressing
2. Press
displayed in the SYS and DIA windows.
3. Press
4. While watching the current meter, press
pressure.
5. Note the highest current reading during inflation.
6. While the pump is running, verify that the reading on the current meter is not more
than 750 mA above the current level noted in
current draw"
7. Press
To test inflation
1. Press
displayed in the SYS and DIA windows.
once to select 80 mmHg (10.7 kPa). The cuff quickly inflates to
to select 150 mmHg (20 kPa) target pressure. The cuff quickly inflates to
to select 300 mmHg (40 kPa). The cuff quickly inflates to approximately 300
Replace the internal battery with an external power supply. (See
substitution cable setup"
WARNING Do not connect the monitor to a patient while the monitor is in
Service mode. Overpressure detection is disabled while the monitor is in NIBP
test mode.
repeatedly until NIBP TEST appears in the message window and 0 is
to select 0 mmHg (0 kPa) target pressure.
on page 13).
three times to select 0 mmHg (0 kPa) target pressure.
repeatedly until NIBP TEST appears in the message window and 0 is
on page 12.)
.
to select 80 mmHg (10.7 kPa) target
step 2
of the verification test
Functional verification
17
"Battery
("Baseline