Checklist and test results report form

Use a copy of this form to track your progress through the functional verification.
Test
Battery charging
Power-on self-test
Beeper
Monitor-off current
Init/idle current
Baseline current
Battery voltage
NIBP characterization test
NIBP verification
Pressure leakage
Pressure accuracy
Valve/pump current
Inflation
Deflation
Max allowable cuff pressure
Overpressure cut-off
Printer current
SpO
Temperature
Nurse call relay
Battery charge time
Battery discharge time
Battery voltage
Monitor serial #:
Tested by:
Test date:
Result
(mA):
(mA):
(mA):
Monitor (V):
Before (mmHg):
80 mmHg
150 mmHg
300 mmHg
Valve/Pump (mA)
- Baseline (mA)
Difference (mA)
Elapsed time (sec):
Manometer reading (mmHg):
Adult
mmHg
Pediatric
mmHg
Neonate
mmHg
(Pass-Fail)
Peak (mA):
Saturation (%):
2
 Fahrenheit:
pins 1-2
pins 2-3
DMM (V):
After (mmHg):
SYS:
SYS:
SYS:
kPa
kPa
kPa
Alarm ()
Alarm ()
volts
Functional verification
Pass
(or
N/A)
Manometer:
Manometer:
Manometer:
Pulse (bpm):
 Celsius:
No Alarm ()
No Alarm ()
31
Fail