Service Procedures; Characterizing Nibp; To Configure To A Different Com Port - Welch Allyn 53000 Service Manual

Vital signs monitor 300 series
Hide thumbs Also See for 53000:
Table of Contents

Advertisement

Service manual

To configure to a different COM port

1. Right-click on the port you want.
2. Select Properties.
3. Select Port Settings.
4. Click Advanced.
5. Select a new COM Port Number.
The port settings for the Keyspan adapter are explained in the documentation included
with the device. Keyspan provides an assistant to configure and diagnose the USB-to-
serial adapter. Access the assistant from Port Settings in the Device Properties dialog.
The VSM uses these default settings:

Service procedures

Characterizing NIBP

This operation trains the NIBP module to properly account for the volumes of internal and
external tubing when the pump is running.
9600 bits per second
8 data bits
no parity
1 stop bit
no flow control
WARNING You must characterize NIBP after replacing any NIBP components:
pump, tubing, air filter, check valve, or NIBP module. Failure to properly
characterize NIBP can lead to injury of neonatal patients.