b. Remove the NIBP module.

NIBP reassembly notes

The shorter tubing section contains the air filter. It connects to the pump input/suction
fitting on the outer edge of the front.
The longer tubing section contains the check valve. It connects to the pump pressure/
output fitting in the center of the pump and to the manifold on the NIBP board.
Air filter

Remove and replace the speaker

1. Perform the following disassembly procedures to access the speaker:
WARNING You must characterize the NIPB module whenever you replace any
NIBP component (pump, tubing, air filter, check valve, or NIBP module).
WARNING If the NIBP module is not properly characterized, the pump could
overinflate a neonatal cuff, which could be hazardous to neonatal patients. (See
"NIBP characterization"
Input
Clear
Caution Do not confuse the NIBP pump power cable connection with the
speaker cable connection. On monitors with the newer NIBP module, the
connectors are similar.
Caution Avoid the following:
Stacking or piling-up unpackaged speakers
Placing the speaker near a magnetic field with high gauss
Rough handling of the speaker assembly
Placing pressure on the speaker magnet and terminal area
Pulling on the speaker wires
Applying direct pressure to the speaker mylar cone
Remove and disconnect the battery.
Separate the front and rear chassis.
Disconnect and remove the main board from the rear chassis assembly.
on page 36.)
Output
White
Check valve
To the NIBP module
Disassembly procedure