If updates are available, Update Status is either "Update required" or "Update
Recommended".
1. Connect the monitor to the AC power adapter and to the PC.
2. Start the utility.
3. Select Update Monitor.
4. To start a software update, click Update. The monitor reboots and proceeds with all
available updates, beginning with the Monitor Boot. During the update process, the
monitor reboots as needed.
The Monitor Boot and Main software modules take about six minutes to load; POEM1
3.20.00 software takes about 15 minutes to load.
Troubleshooting and repair
47