Temperature Module Reassembly Notes; Disassemble The Nellcor Spo 2 Assembly - Welch Allyn 53000 Service Manual

Vital signs monitor 300 series
Hide thumbs Also See for 53000:
Table of Contents

Advertisement

82
Disassembly procedure

Temperature module reassembly notes

To install the temperature board in the module housing:
1. Square the circuit board in the housing to fit the probe-well receiver into the top of the
outside housing.
2. Tilt the board and rotate it counterclockwise so that the temperature cable connector
(J5) fits down into the connector slot in the housing.
Disassemble the Nellcor SpO
1. Remove the SpO
a. Remove the four screws (located on the inside of the rear chassis on the right
b. Remove the side panel and board assembly form the rear chassis.
assembly from the rear chassis and disassemble it as follows:
2
side) securing the SpO
2
assembly
2
assembly to the rear chassis.
Four screws securing the SpO
assembly to the rear chassis
Welch Allyn VSM 300 Series
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents