24
Functional verification
Testing the sensor with the monitor
Use this procedure to test the monitor SpO
1. Connect an SpO
2. Set the BioTek Index2 SpO
SpO
% SpO
Rate
3. Wait for the monitor display to stabilize.
4. Verify an SpO
5. Verify that the monitor displays a pulse rate of 60 ± 4.
sensor to the simulator optical finger and to the monitor.
2
2
manufacturer type
2
2
saturation level of 94 ± 4.
2
function while verifying a sensor.
2
simulator as follows:
Nellcor (MP-506A or Nell-3A)
Masimo (MS-11)
94
60 (Motion Artifact not selected)
Welch Allyn VSM 300 Series