2
Installation

Removing the Existing Foreline Pump from the Agilent 7000 and 7010 Series GC/MS

Removing the Existing Foreline Pump from
the Agilent 7000 and 7010 Series GC/MS
The existing foreline pump may be hot. The foreline pump can cause burns if
WA R N I N G
touched when operating.
To remove the existing foreline pump from the GC/MS:
1 Vent and shut down the instrument. Venting begins the automated shut
2 Turn off the GC/MS.
3 Disconnect the instrument from its power outlet.
4 On the GC/MS foreline gauge adapter, loosen the hose clamp, and carefully
12
down of the GC/MS. Refer to the Agilent 7000/7010 Series Triple Quadrupole
Operating Manual for details.
remove the existing foreline hose from the end of the foreline gauge adapter
as shown in
Figure
2.
Figure 2.
Foreline hose removal from foreline gauge adapter
Agilent G6697A Oil Free IDP-10 Dry Scroll Pump Installation Guide