Measurement Mode Auto-V; Applications; Choosing The Probe; Activating Auto-V - GE DMS Go+ Operating Manual

Table of Contents

Advertisement

7 Special functions of the DMS Go+ TC

7.2 Measurement mode Auto-V

Applications

The Auto-V function is used for the simultaneous deter-
mination of sound velocity and wall thickness on non-
coated test objects.
The simultaneous determination of sound velocity is not
possible in wall thickness measurements on coated test
objects.
A T T E N T I O N
A prerequisite for correct measurement re-
sults is a good homogeneity of the material in
the immediate vicinity of the measurement
location.
As opposed to that, large-area inhomogene-
ities have an effect on the calculation of wall
thickness.
7-6

Choosing the probe

To be able to use the Auto-V function, you have to con-
nect a special 4-element probe to the DMS Go+.
The following probes from GE are suitable for Auto-V:
● TC-560

Activating Auto-V

– Connect a suitable probe to the instrument.
– In the function group PROBE&CAL, select the func-
tion PROBE.
– Choose the connected probe. The instrument is im-
mediately adjusted.
– Select the function TG MODE and choose the option
AUTO-V.