5 Operation
Gate width
You can set the width of gate A and gate B.
– In the function group GATE, select the function
GATE A WIDTH.
– Set the required width for gate A.
– If the selected measurement mode allows it, select
the function GATE B WIDTH and set the required
width for gate B.
5-38
Gate threshold
You can set the vertical position of the gate in percent-
age of screen height (A-scan).
N o t e
In RF mode, the values between +100% and
–100% are possible. When changing from
RF to NEG, POS, or FULL, negative signs
are replaced by positive signs (see Section
Selecting the rectification mode, page 5-
36).
– In the function group GATE, select the function
GATE A THR.
– Set the required position of gate A.
– If the selected measurement mode allows it, select
the function GATE B THR. and set the required posi-
tion of gate B.
Edition 4 (05/2014)
Setting the gates
DMS Go+