5 Operation
– In the function group PROBE&CAL, select the func-
tion TEMP COMP.
– Press one of the left or right arrow keys in order to
choose the option ENABLE.
– Select the function CAL TEMP and set the tempera-
ture of the calibration block.
– Select the function CUR TEMP and set the tempera-
ture of the test object.
– Select the function TEMP COEF and set the tem-
perature coefficient.
– Finally, press the function key 1 (function HOME) in
order to exit the function. The settings become effec-
tive immediately.
N o t e
The DMS Go+ can display both the uncom-
pensated and the compensated thickness
value at the same time. To achieve this, con-
figure the data boxes accordingly (see Sec-
tion Configuring data boxes, page 5-13).
5-52

Display freeze (FREEZE)

You can freeze the display screen of the DMS Go+. The
A-scan and the displayed thickness value are then fro-
zen and no longer updated.
This function is available in all measurement modes and
display views.
When the function is active, a corresponding icon is dis-
played at the top left corner of the display screen (see
Section Status display icons, page 0-5).
– In the function group MEASURE, select the function
FREEZE.
– Press one of the left or right arrow keys in order to
switch the function on.
N o t e
You can assign the function FREEZE to the
function key 4 (see Section Assignment of
function key 4, page 5-7).
Edition 4 (05/2014)
Other functions
DMS Go+