Configuring alarm functions
5.10 Configuring alarm functions
The DMS Go+ can trigger an alarm signal if preset read-
ings are exceeded or fallen below.
If an activated alarm is triggered, the displayed value on
the screen flashes red and the red alarm icon flashes in
the status display box (see Section Status display
icons, page 0-5).
DMS Go+
Minimum value alarm
The alarm is triggered if the set minimum value is fallen
below.
– In the function group ALARM, select the function
MIN.
– Choose the option ON in order to activate the mini-
mum value alarm.
– Select the function MIN ALARM and set the thick-
ness value which should trigger an alarm if fallen be-
low.
Maximum value alarm
The alarm is triggered if the set maximum value is ex-
ceeded.
– In the function group ALARM, select the function
MAX.
– Choose the option ON in order to activate the maxi-
mum value alarm.
– Select the function MAX ALARM and set the thick-
ness value which should trigger an alarm if exceeded.
Edition 4 (05/2014)
5 Operation
5-39
Need help?
Do you have a question about the DMS Go+ and is the answer not in the manual?
Questions and answers