108
Speech Intelligibility
STI-PA results screens
CAP-1
Press first
softkey to assign
a captured curve
name (i.e., CAP-
1, CAP-2, CAP-
3, or CAP-4)
5. Once all background noise has been captured, refer to, "Applying captured curves"
below.
Applying captured curves for STI-PA testing
To apply the curves during a STI-PA measurement remember to complete one
and two below before applying the curve.
Applying captured curves
1. From the Start screen, select View Session and press
2. Select one of the captured curves by pressing the 1
3. From the Start screen, press the 1
Figure 6-26: Captured curves/background noise
Press On/Off/Esc key.
selected. Then, select View Session and the STI-PA screen will open.
Figure 6-27: Post CAP-1 applied during STI-PA test
Curve indicator
Capture
Press Capture
softkey to save
the background
noise/SPL.
st
softkey (i.e, Cap-1).
softkey continuously until STI-PA is
st
POST CAP-1
The selected Captured Curve,
i.e., CAP-1, will appear on the
STI-PA screen. To change the
Cap-x, repeat steps 1-2 above.
.. Press Altf key.
Models SE and DL