OMGEM5KN4.7.1
A full charge will take approximately 4 hours from a fully discharged battery.
Warning
The battery charger is intelligent and will indicate when the unit is charging and charged.
The instrument must be charged ONLY using the battery charger supplied with the
instrument. The battery charger supplied is intended for indoor use only. Please ensure
adequate ventilation while charging. Typically, a fully charged battery will last 7-8 hours.
A quick 30 minute charge can be used to give approximately one hours use in the field but
this may shorten the battery life. Temperature can dramatically affect the battery life;
please take this into account when estimating battery life.
Note:
Connect the charger to the mains attaching the appropriate adaptor.
Power supply front and back drawing:
Charger:
Input voltage:
Input frequency:
Input current:
Output voltage:
Output current:
Note:
This charger has been internally restricted to 1.5A

6.10 Cleaning instructions

Do NOT use any cleaning agents to clean the analyzer or battery charger as they may
have an adverse effect on the safe use of these devices.
Page 20 of 70
The battery charger is NOT covered by the Ex certification.
The battery must be charged only in a safe area.
100-240V AC +/- 10%
50-60Hz +/- 10%
0.4A@100VAC .. 0.2A@240VAC
10.1VDC max
1.5A max
GEM5000 Gas Analyzer