LANDTEC GEM5000 Operating Manual page 46

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

OMGEM5KN4.7.1
Instructions for use:
1)
Attach the anemometer to 'Connector C' (refer to section 5.3 – Instrument
connection points).
2)
Place the anemometer into the pipe (sample point) ready to take the reading.
3)
To take a flow measurement when using an anemometer, follow the instructions
displayed on the analyzer. When the reading has stabilized press the 'Enter' key to
store the reading.
Anemometer cleaning instructions:
General handling tips:
Protect the probes against severe vibration.
Do not kink the connector cable (risk of cable breakage).
Never allow hard objects to contact rotating impellers.
Always carry out probe cleaning according to the cleaning instructions.
Never immerse probes in solvent.
Never blow probes through with compressed air.
Allow hot probes to cool slowly, never cool by plunging them in cold water etc.
Cleaning instructions:
Instrument and probe must be switched off or disconnected prior to cleaning.
Vane probes:
As the probes are highly sensitive measuring instruments, they must be cleaned with
great care.
Fibres or other foreign bodies can be carefully removed with fine tweezers. When
doing so, take care not to bend or otherwise damage the vanes or the spindle.
The adjustment of the bearing screws must never be changed. This can result in an
erroneous measurement.
Never allow hard objects to contact rotating impellers.
Cleaning agents that extract the plasticizer from the plastic are never to be used for
plastic probes (practically all solvents).
Cleaning the probes – best practice:
Note:
Use soapy water.
Cleaning example:
1)
Carefully, swish the top part of the impeller back and forth in clean soapy water for
approximately 10 minutes. Then swish the top part of the impeller back and forth in
clean soapy water. If soapy water is used as a cleaning agent it is advisable to wash
out the soap solution thoroughly with distilled water.
2)
After cleaning the probe, rub it dry with a clean, dry cloth.
Page 46 of 70
GEM5000 Gas Analyzer