Device Settings; Date And Time - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 Gas Analyzer
4) The operator will return to the 'User settings' menu.
Note:
If no technicians are loaded via LSGAM, this section is skipped during
start up and the 'Technician ID' icon is removed from the menu.

7.3 Device settings

To access the 'Device Settings' menu, select the 'Menu' key on the front of the analyzer to
display the 'Operating Settings' menu followed by the soft-key to display 'Device Settings'
menu. The following menu is displayed:

7.3.1 Date and time

This option enables the operator to set the instrument date and time or to
receive and update the settings automatically from satellite signal.
1)
Select the 'Menu' key on the front of the analyzer to display the 'Device Settings'
menu followed by 'Key 1 – Date and Time' and the following screen is displayed:
2)
Select 'Key 1 – Set Date' and key in the required date. Type the date using the
numeric keypad. Press the soft-key 'Date Format' to toggle and select the required
date format i.e. dd/mm/yy. Press the 'Enter' key to confirm and update the date
setting.
3)
Select 'Key 2 – Set Time' and key in the required time (hh:mm). Type the time
using the numeric keypad and press the 'Enter' key to confirm the update.
OMGEM5KN4.7.1
Page 27 of 70

Advertisement

Table of Contents
loading

Table of Contents