Note:
When onsite the site engineer must have an unmodified water trap assembly
in addition to the modified water trap with a H
readings with and without a filter.
1)
Once the H
analyser displayed when taking readings using a H
Note:
Fitting a pitot tube to a sample point when using an H
2)
Make sure the water trap filter is fitted as close as possible to the pitot tube. This
will help protect and trap moisture before it reaches the H
the water trap filter is clean and dry.
3)
Attach the sample tubing into the analyzer.
4)
To take flow readings using a pitot tube, please follow the instructions displayed on
the front of the instrument. The user will be prompted to complete each step and
should follow the instructions on screen.
Note:
Page 50 of 70
S filter is fitted follow the instructions as detailed on the front of the gas
2
This is only for analysers fitted with the CO compensated cell.
It is important to fit the pitot tube central and parallel with the pipe.
GEM5000 Gas Analyzer
S filter in order to take
2
S filter.
2
S filter:
2
S filter. Make sure that
2