Calibration - Mixtures 1, 2, & 3 - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 Gas Analyzer
9.7.2
Calibration – mixtures 1, 2, & 3
Switch on the
instrument and
START
complete the
instrument warm up.
Connect the flow
regulator to the
white port on the
analyser
Flow certified gas
mixture using the
gas regulator
supplied
Was the
calibration
successful?
Unsuccessful
Please see section
9.11 – Calibration
troubleshooting
Main Gas Read
Screen select the
'Menu' key
Press
'Enter'
to confirm
Gas mixture
Wait for the readings
Select
to stabilise. Wait for
soft-key
the analyser to
'Check'
check spans for
approx. 1 minute.
Read status field
in the right hand
corner of the
screen
Retry
Calibrate
Calibrate
Record
Successful
Read status field
in the right hand
corner of the
screen
Select
soft-key
'Exit'
END
Select correct gas mixture:
1, 2, or 3, or add User-
selectable gas mixture
Message will
display 'Please
wait for timer to
reach zero'
Record
OMGEM5KN4.7.1
Select key 2
'Gas check '
Select key 1
"Gas check"
When complete a
message will
display 'Check
complete'
This feature records
that a gas check has
been completed but
does not calibrate the
analyser
Select
soft-key
'Exit'
Page 61 of 70