Adjust Flow Fail - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 Gas Analyzer
2) Select 'Key 4 – Set Alarms' and the following menu is displayed:
3) Select the corresponding key to select the gas for which you wish to set an
alarm/target trigger for, followed by 'Key 1' to change the trigger condition of an
alarm.
4) To manually adjust the alarm/target set press (<) 'Key 4 – Scroll left' or 'Key 6 –
Scroll right' (>) and enter the trigger value. Once you are happy, press the middle soft
key for 'save'.
5) For pressure, temperature and flow alarms, press the left soft key for 'Secondary' and
then select the corresponding key to select the channel for which you wish to set an
alarm trigger for, followed by 'Key 1' to change the trigger condition of an
alarm/target. Once you are happy, press the middle soft key for 'save'.
6) To disable all alarm settings select key 0 – 'Disable All'
Note:

7.2.5 Adjust flow fail

This option enables the operator to adjust the flow fail tolerance of the
instrument, i.e. the operator can adjust the sensitivity for when the pump
will stop operating on the presence of a blockage or low flow.
1) Select the 'Menu' button on the front of the analyzer to display the 'Device Settings'
menu. Press the soft-key to display 'Operation Settings'.
ID specific alarms cannot be added/edited on the analyzer, to add/edit
ID specific alarms, please use the optional LSGAM Software.
OMGEM5KN4.7.1
Page 25 of 70

Advertisement

Table of Contents
loading

Table of Contents