Alarms; Settings - LANDTEC Geotech G100 Series Operating Manual

Table of Contents

Advertisement

G100 Range
5.9.2.3

Alarms

The G100 range of instruments has the facility to set rising or falling alarms for the two
main gas channels, CO
can be enabled or disabled independently via the 'Alarms' menu option.
Once enabled these alarms become active in the 'Main Read Screen'; this is indicated by
a bell
icon. If an alarm is triggered the screen turns red and a flashing bell
displayed. The beeper is sounded until the gas level has recovered beyond the trigger
point.
Rising alarms are triggered when the gas level exceeds the maximum value entered
by the user.
Falling alarms are triggered when the gas level falls below the minimum value
entered by the user.
1)
From the 'Main Read Screen' press the 'Menu' key on the instrument panel.
2)
Press 'Key 2' to display the 'Utilities' menu.
3)
Press 'Key 3' to select alarm settings and the following screen is displayed:
4)
Press 'Key 1' to maintain CO
5)
Select from the following:
1-Enabled/Disabled
2-Max
3-Min
6)
Select the option to modify followed by the soft-key 'Accept' or 'Reject'.
5.9.2.4

Settings

The 'Settings' option enables the operator to maintain information with regards to
taking samples and readings.
1)
From the 'Main Read Screen' press the 'Menu' key on the instrument panel.
2)
Press 'Key 2' to display the 'Utilities' menu.
3)
Press 'Key 4' to select settings and the following screen is displayed:
 Copyright Geotechnical Instruments (UK) Limited
and O
(if selected as an option). The alarms for each channel
2
2
Current Alarms
alarm and 'Key 2' to maintain O
2
Toggle between disabled and enabled alarm status.
Sets the upper limit alarm setting.
Sets the minimum alarm setting. The default is 0.
OMG100N1.21
icon is
alarm.
2
Page 19