PNA Series Microwave Network Analyzers
Tests and Adjustments
N5221A/22A
ANSI/NCSL Z540.3–2006 and ISO/IEC 17025 Verification
ANSI/NCSL Z540.3–2006 and ISO/IEC 17025 Verification
To meet the criteria for ANSI/NCSL Z540.3–2006 and ISO/IEC 17025 verification, perform the preliminary
1
checks and all performance tests without stopping to repair or adjust
. Refer to
Figure 3-1
for test flow. Print
data at the completion of all the tests, even if you are aware that the analyzer did not pass. If there is a
failure, complete the verification before you troubleshoot, repair, and adjust. After the failure has been
corrected, repeat the entire set of performance tests and generate a new set of data.
Figure 3-1
ANSI/NCSL Z5240.3–2006 and ISO/IEC 17025 Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage
Service Guide N5222-90001
3-9