PNA Series Microwave Network Analyzers
Tests and Adjustments
N5221A/22A
Adjustments
Figure 3-18
PNA Noise Adjustment Dialog Box
3. Click Begin.
4. As shown in
Figure
3-19, connect an RF cable between the network analyzer test ports 1 and 2.
Figure 3-19
Equipment Setup 1 for the Receiver Adjustment
5. When prompted, connect the noise source to port 2, either directly or with a short, low-loss adapter. See
Figure 3-20
Service Guide N5222-90001
3-49