Keysight N5221A Service Manual page 24

2-port and 4-port pna microwave network analyzers (10 mhz - 13.5 ghz) (10 mhz - 26.5 ghz)
Table of Contents

Advertisement

General Product Information
Required Service Test Equipment
Required Service Test Equipment (Cont'd)
a
Equipment
3.5 mm calibration kit
3.5 mm verification kit
BNC cable (2 required)
3.5 mm RF cable (Qty 2)
GPIB cable
3.5 mm (f) to 3.5 mm (f)
3.5 mm (f) to type-N (m)
3.5 mm (m,f), 10-dB fixed
attenuator
3.5 mm (m,f), 20-dB fixed
attenuator
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, R = Repair, V = System verification
c. Included in the 85052B/D calibration kits.
2-6
Critical Specifications
Calibration and Verification Kits
Freq: 10 MHz to 26.5 GHz
Freq: 10 MHz to 26.5 GHz
Cables
50 , length  60 cm
50 , length  60 cm
N/A
Adapters
Return Loss:  32 dB
Return Loss:  28 dB
Attenuators
Accuracy: ± 0.5 dB
Freq: 10 MHz to 26.5 GHz
Accuracy: ± 0.5 dB
Freq: 10 MHz to 26.5 GHz
PNA Series Microwave Network Analyzers
Recommended
Alternate Model
Model or Part
or Part Number
Number
85052B
85052D
85053B
None
8120-1839
None
85131C
85131E
10833A/B/C/D
None
83059B
85052-60012
1250-1743
None
8493C
None
Option 010
8493C
None
Option 020
Service Guide N5222-90001
N5221A/22A
b
Use
P,T,V
V
A
P,A,V
P,A
c
P,A,T
P,A,T
P
P

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5222a