Tests and Adjustments
PNA Series Microwave Network Analyzers
Non-Standards Compliant Verification
N5221A/22A
Non-Standards Compliant Verification
To meet the criteria for non-standards compliant verification, perform the preliminary checks and the
performance tests while stopping to troubleshoot. Refer to
Figure 3-2
for test flow. Troubleshoot and repair
the first problem encountered without continuing to other tests. After you troubleshoot, repair, and adjust,
repeat the last failed portion and generate a new set of data.
Figure 3-2
Non–Standards Compliant Verification Flowchart
3-10