Troubleshooting
Rear Panel Troubleshooting
Table 4-3
LAN Pin Definitions and Wire Color Codes
Pin Number
Color
1 (transmit +)
White/orange
2 (transmit )
Orange
3 (receive +)
White/green
4
Blue
2. Cut the wires going to pins 1, 2, 3, and 6. Strip away a small amount of insulation from each of the eight
cut ends.
a. Connect the wire from pin 1 on one end of the cable to the wire from pin 3 on the other end of the
cable.
b. Connect the wire from pin 3 on one end of the cable to the wire from pin 1 on the other end of the
cable.
c. Connect the wire from pin 2 on one end of the cable to the wire from pin 6 on the 
other end of the cable.
d. Connect the wire from pin 6 on one end of the cable to the wire from pin 2 on the 
other end of the cable.
3. Insulate all exposed wires so that they cannot short together.
4. Label this as a crossover cable so that it cannot be confused with a standard cable.
Figure 4-5
Construction of a Crossover Cable
4-22
PNA Series Microwave Network Analyzers
Pin Number
Color
5
White/blue
6 (receive )
Green
7
White/brown
8
Brown