7 .22 Other functions

7 .22 .1

Switching of the wireless

communication function
of the DR Detector
The wireless communication functions of the DR
Detector (permitted/prohibited) can be switched .
IMPORTANT
••••• ••••••••••••••••••••••••• •••••••
• Switch the wireless communications function from off
to on while the DR Detector is in the wired connection
condition .
• If there are more DR Detectors than I/F Cables, estab-
lish a wired connection once for DR Detectors that did
not have one when the wireless communications func-
tion was switched from off to on before use.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
1
Press [System] on the initial screen .
• The system monitor screen is displayed .
• If the network status is not displayed, select the
[System Monitor] tab on the function switching tabs.
2
Press [Wireless] to switch the wireless
communication function .
• [
] shows that wireless communica-
tion function is on .
• [
] shows that wireless communica-
tion function is off.
• A confirmation screen is displayed.
3
Press [OK] .
• Press [Cancel] to return to the previous screen.
• The wireless communication function switches .
4
Press [Close] .
• The display returns to the initial screen .