8.2 Dialog boxes/balloons and countermeasures
Displayed message
Before study completion, please output or reject each Mammo
Image .
Below consoles are having more studies, but only partial stud-
ies are displayed .
Birth date is mandatory, Please enter birth date .
Calibration of [DR Detector name] have expired.
Please perform calibration urgently .
Do it now?
- If choose later:
Press the arrow buttons at the bottom of the screen .
And Press the calibration button in the toolbar that appears .
Calibration of [DR Detector name] will expire in [LimitDate] more
day(s) .
Calibration of [DR Detector name] will expire in [LimitDate] more
day(s) .
Please perform calibration .
Do it now?
- If choose later:
Press the arrow buttons at the bottom of the screen .
And Press the calibration button in the toolbar that appears .
Can not manipulate the data in the deleted status .
Cannot add exposures beyond [0].
Cannot add exposures beyond [MaxSize].
Cannot capture images due to low battery power .
Charge the battery .
Cannot edit the stitching position, because image stitching is
performed at other console . Please stitch the images again .
Cannot get exposures beyond [0].
Cannot perform the calibration in the Technologist .
Please perform again in higher than super user privileges .
Cannot perform the operation as selected study data save is in
progress .
Please perform the same operation after some time .
Cannot perform the studies as different Patients studies got
selected .
Cannot perform the study as the study is in moving .
Cannot perform this operation as image reading is inprogress
Problems/countermeasures
This is displayed when finishing an examination while there are
mammogram images that have not yet been output .
Output all mammogram images or reject them .
Displayed when unable to acquire parts of a study after per-
forming multiple terminal study acquisition through study shar-
ing . Check the connection and reacquire the study .
Displayed when mandatory fields have not been filled in.
Press [OK] to close the screen.
This is displayed when calibration of the DR Detector has ex-
pired .
Perform calibration as soon as possible .
The calibration expiration date of the DR Detector is displayed.
This is displayed when calibration of the DR Detector will be
expiring soon.
Perform calibration .
The study cannot be opened as it has been deleted .
Displayed when the number of scheduled exposures for the
same patient exceeds the maximum value. Further reservations
cannot be added .
This is displayed when the maximum number of Exam. Tags
that can be registered has been exceeded.
Displayed when the battery level for the DR Detector in use is
low . Charge the DR Detector .
Displayed when trying to recombine an image that was com-
bined on another shared terminal . Combine the images again
on the original shared terminal .
The Image Count for this patient has exceeded the upper limit.
No more studies can be acquired .
This is displayed when you log in as Technologist and try to
perform calibration .
Log in again using Super user or higher privileges .
The selected study was displayed during image processing .
Wait until processing is finished and perform the operation
again .
Displayed when a different study is performed with a multi-
study . Reselect a study of the same patient .
Displayed when the [Perform] button is pressed while a study is
being transmitted or the destination device is receiving a study .
Cancel sending the transmission or start the study with the
destination device .
Displayed when attempting to shut down the system or delete
a job while reading an image . Shut down the system after the
image reading is finished.