7.2 Exposure with AeroDR Portable RF Unit2
• The selected Exam. Tag is added to the Exam.
Tags display area . By repeating from step 1,
Exam. Tags can be added.
4
When finished selecting/adding the Exam.
Tags, press [Perform] .
• The exposure screen is displayed.
HINT
•••••••••••••••••••••••••••••• •••••••
• When only reserving an examination, after adding the
Exam. Tag, press [Schedule]. After reservation is per-
formed, the initial screen is displayed .
• Returning to the initial screen by pressing [Schedule]
does not set study date and time of image supplemen-
tal information automatically .
• It is also possible to register multiple exposure methods
in 1 Exam. Tag. Register these methods when install-
ing .
• If you have set to register the Exam. Tags automati-
cally, the body part selection screen is not displayed .
The display transitions to the exposure screen.
• •• • • •• • • •• • •••••••••••••••••••••••••••••••••••••••••
7 .2 .5

Exposure

Prepare for exposure with the AeroDR Portable RF
Unit2 according to the following procedure .
IMPORTANT
• Always check the patient information and exposure
conditions before exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
1
Confirm that each device is ready for ex-
posure .
IMPORTANT
• Check the wireless connection .
• If a wired connection is established between this de-
vice and the AeroDR Portable RF Unit2, disconnect
the Ethernet cable that connects this device and the
hub .
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
2
Perform exposure .
• Configure the mobile X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
• The exposed image is displayed in the image
display area .
IMPORTANT
• When performing an exposure, stay away from the pa-
tient as far as possible before pressing the exposure
switch .
• Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
• If an image is displayed when an exposure is not per-
formed, this may be a detection error due to noises,
etc . Perform rejected image (deletion of images) and
repeat exposure.
• If you have performed an exposure when "Ready" was
not displayed, repeat an exposure after approximately
one minute .
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD . Selecting the AeroDR 3 1417HD
icon allows the exposure time to be selected.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••