Flow Cell Installation - Hach QC8500 Series 2 User Manual

Fia automated ion analyzer
Table of Contents

Advertisement

Installation
1
On/Off switch
2
Power for system unit
3
High pressure pump (IC only)

3.1.4 Flow cell installation

1
To waste
2
Flow cell body
12
5
4
1
3
2
Figure 4 System unit connection
To install the flow cell:
1. Insert the flow cell into the detector with the printed side (bottom) down. Since the
flow cell is keyed, insert the flow cell only in the corresponding slot of the detector
(Figure
5).
2. Make sure the flow cell is pushed all the way in the slot. The bottom flared tubing
connector on the flow cell is the inlet and the top connector is the outlet. During the
installation of a manifold (reaction module), it is very important to make the
connections accordingly, so air bubbles do not get trapped in the flow cell. Refer to
Figure 6 on page
13.
3. Connect the top connector (outlet) to the waste line.
Figure 5 Flow cell
4
Pump connectors
5
USB in (from computer)
3
Inlet
1
2

Advertisement

Table of Contents
loading