Hach QC8500 Series 2 User Manual page 38

Fia automated ion analyzer
Table of Contents

Advertisement

Operation
4.2.1.6 Tee connector symbol
1
Tee Connector
4.2.1.7 Heating coil symbol
36
The reagent lines are labeled to indicate the solution going through that line. The carrier
line is always connected to port 2 of the valve. The sample line is always connected to
port 6 of the valve.
Note: The manifold diagram examples have only one tee connector.
Every intersection of two reagent lines represents a tee connector
Tee connectors combine two reagents. The tee has three ports; two side ports (inlets)
where the two reagent lines are connected and a top port (outlet) which may be
connected to a mixing coil or whatever is indicated in the manifold diagram.
Figure 35 Tee connectors symbols
Figure 36
represents a heating coil installed on the System Unit. Set the heater at the
temperature specified in the diagram (60° C in the example). The length of tubing
wrapped on the block is specified in the QuikChem method.
Figure 36 Heating coil symbols
(Figure
35).