Operation

4.9 Pump operation

Button
MIN
MAX
NORMAL RUN
MANUAL RUN / STOP
arrow
UP
arrow
DOWN
46
To use new pump tubes for the first time, allow the pump to run approximately ten
minutes in order to obtain constant flow rates.
on the pump.
Table 8 RP-150 pump button functions
Function
Sets the pump speed to 4.
Sets speed to maximum of 999. Button must be kept depressed in order to maintain
maximum speed.
Note: If this button is depressed too long the pump will stop to prevent the motor burning out.
Normal run starts the pump and permits remote control of the pump standby speed as determined
by the Omnion software at the end of a run or batch. The pump runs at the displayed speed.
Normal speed is 35. Use the
If Omnion is set to Standby speed, the pump will run very slowly and the display will show rc 3.
The display rc 3 indicates remote control speed 3.
Manual running allows the pump to override the remote control of the standby speed.
This button toggles between manual speed control and manual stop.
If the pump is running, this button will always stop the pump. If the pump is not running, this button
will still override the remote control from the computer and put the pump in manual stop.
The
button is commonly used for making the pump go to normal speed in order to
MANUAL RUN
prime the tubing with reagents.
Increases pump speed setting. Not active when in
Decreases pump speed setting. Not active when in
Place all of the reagent lines into their corresponding containers or in deionized water. To
avoid reagents leaking on the system, run deionized water through all of the lines to
detect any leaks. If the manifold has a cadmium column, make sure it is bypassed. Water
and air will damage the cadmium column.
Note: Some methods require certain reagents to be pumped first. For example, the TKN method
has a buffer solution that has to be run first to avoid a white precipitate in the manifold tubing. Check
the QuikChem method for instructions.
Refer to
section 3.2.3 on page 18
reagent pump.
Table 8
and
arrows to alter this setting.
UP
DOWN
or
MIN
MIN
to learn how to mount the cartridges on the
shows the functions of each button
modes.
MAX
or
modes.
MAX