Hach QC8500 Series 2 User Manual page 9

Fia automated ion analyzer
Table of Contents

Advertisement

Module
PDS200
ASX-500 sampler
ASX-400 sampler
Reagent pump
System unit
Temperature
Use
Altitude
Relative humidity
Pollution degree
Installation category
Certification Marks
CE
cETL
ETLus
Table 2 Individual electrical specifications
AC input voltage
AC input frequency
rating or range
rating or range
110–240 VAC
100–240 VAC
100–120 VAC
100 or 120 VAC
220 or 240 VAC
100–115 V ± 10%
230 V ± 10%
Table 3 Environmental specifications
Table 4 Certifications
Standards
IEC/EN 61010-1 per LV Directive (2006/95/EC), IEC/EN 61326 per EMC: Directive
(2004/108/EC
CSA C22.2 No 61010-1
UL 61010-1
Maximum active
power (W),
apparent power
(VA) or current
rating
50–60 Hz
0.5 A
50–60 Hz
1.5 A
50–60 Hz
200 VA
50–60 Hz
15 VA
50–60 Hz
15 VA
50–60 Hz
800 W
50–60 Hz
800 W
15–35 °C (60–95 °F)
Laboratory
Up to 10,000 feet (3 km) above sea level
Up to 92.5% non-condensing
II
II
Specifications
Fuse rating
(A)
T, ½ A, 250 V
Fuse not
user-replaceable
Fuse not
user-replaceable
T, ½ A, 250 V
T, ¼ A, 250 V
T, 8 A, 250 V
T, 4 A, 250 V