Pds200 Connections - Hach QC8500 Series 2 User Manual

Fia automated ion analyzer
Table of Contents

Advertisement

Installation

3.4.2 PDS200 connections

4
1
Ferrule
2
Flush
3
To pump/analyzer port B
26
Note: Probe speed on existing autosamplers has been reduced to improve performance. The
autosampler may produce a different sound as a result.
To connect the PDS200:
1. Connect the DI Water Line to Port A of the solenoid valve of the dilutor. Place the
other end of the line into the DI water
2. Connect the Sample Line Assembly to port B of the solenoid valve
(Figure
23, item 3).
3. If necessary, attach the Dual Probe Assembly to the autosampler. The probe guide
may need to be rotated to accommodate the dual probe.
4. Connect the probe sample line (attached to the bent tube of the Dual Probe
Assembly) to port C of the solenoid valve
5. Make sure the ferrule on the line from the straight tube of the Dual Probe Assembly is
flush with the end of the tubing
6. Insert the connector from the straight tube of the Dual Probe Assembly into port D of
the dilutor
(Figure 24 on page
7. Make sure the ferrule on the dilutor end of the Diluent Line Filter Assembly is flush
with the end of the tubing
8. Insert the connector from the Diluent Line Filter Assembly into port E of the dilutor
and tighten
(Figure
23, item 3).
9. Insert the filter end of the Diluent Line Filter Assembly into the diluent
item 2).
3
Figure 23 Tubing connectors and ferrule placement
(Figure
(Figure
(Figure
23, item 2).
27) and tighten.
(Figure
23, item 2).
5
4
To DI water port A
5
To sample probe line port C
23).
23, item 5).
(Figure
23,
1
2