B.3 Manifold installation
1
Reagent pump
2
Pump flow
3
To probe wash (green)
4
Regenerant (red)
5
DI water (gray)
6
To waste
7
To port 6 of next valve or waster
8
Eluent pump
Most IC manifolds will have two columns: a guard column and an analytical column. The
purpose of the guard column is to protect the analytical column. Any unwanted
particulates will be retained in the guard column. Some organic compounds, oils and
grease that may damage the packing material of the columns would be trapped in the
guard column before getting to the analytical column.
manifold diagram. Refer to the QuikChem method for specific installation requirements.
Figure 70 QuikChem manifold diagram for IC (example)
Figure 70
9
Pump flow
10 Sample (orange)
G—Guard column
A—Guard column
S—Guard column
CM—Guard column
V1—Six port valve
V2—Ten port valve
Ion Chromatography