Installation
3.1.5.2 Waste line
1
13 cm Teflon 0.8 mm ID
2
Adapter
3
150 cm Tygon 1.5 mm ID
4
Glass weight
14
A waste line consists of a 13 cm of Teflon tubing 0.8 mm ID, a pump tube adapter, 150
cm of Tygon tubing 1.5 mm ID, a glass weight and a plastic nipple. Refer to
Figure 9 Waste line
To install a waste line:
1. Connect a waste line to port 5 of the injection valve and the top connector of the
flow cell.
2. When running two or more channels simultaneously, interconnect adjacent valves by
installing 15 cm of 0.8 mm Teflon tubing between ports 5 and 6
page
15). The waste channel farthest from the System Unit will drain directly into the
waste container. Refer to
5
To waste container
6
Nipple
7
From flow cell or valve (port 5)
Section 4 on page 29