Welch Allyn CP 100 Directions For Use Manual

Welch Allyn CP 100 Directions For Use Manual

12-lead resting electrocardiograph
Hide thumbs Also See for CP 100:
Table of Contents

Advertisement

CP 100
12-Lead Resting
Electrocardiograph
Directions for Use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CP 100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Welch Allyn CP 100

  • Page 1 ™ CP 100 12-Lead Resting Electrocardiograph Directions for Use...
  • Page 2 Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician. Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
  • Page 3: Table Of Contents

    Contents 1 - Introduction ..........1 About This Manual .
  • Page 4 Contents Welch Allyn CP 100 Electrocardiograph 4 - Reviewing the ECG Settings....... 31 “ECG Settings”...
  • Page 5: Introduction

    Introduction About This Manual ..........2 Product Overview .
  • Page 6: About This Manual

    Product Overview The Welch Allyn CP 100 electrocardiograph features a full alphanumeric keyboard, an LCD display, full-size user-programmable reports, and the ability to operate on either battery or AC power.
  • Page 7: Indications For Use

    Removable leads for ECG patient cable Replace leads individually if needed. Compatibility with CardioPerfect workstation software Store and manage data electronically by transferring records to a Welch Allyn CardioPerfect workstation via an SD memory card or removable USB storage device (media not included).
  • Page 8 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph Pacemaker detection If the software detects the possible presence of a pacemaker, it asks you whether the patient has a pacemaker. If you say no, interpretation (if purchased) is included in the report.
  • Page 9: Options

    Directions for Use Chapter 1 Introduction Options These options are available both for initial purchases and for upgrades. • Automatic ECG interpretation The optional MEANS interpretation algorithm, developed by the University of Rotterdam in the Netherlands, provides automatic analysis of ECG tests. For more information, see the MEANS Physicians' Manual or the PEDMEANS Physicians’...
  • Page 10: Accessories

    Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph Accessories To order accessories, call Welch Allyn. For phone numbers, see page Item Customer Order Number Quantity Resting tab electrodes 45008-0000 1000 Resting tab electrode adaptors 58581-0000 Thermal chart paper (1 case = 5 pads, 200 sheets each)
  • Page 11: Controls, Indicators, And Connectors

    Directions for Use Chapter 1 Introduction Controls, Indicators, and Connectors This section describes the controls, indicators, and connectors that are part of the electrocardiograph. Figure 3. Top Keyboard Figure 1 on page 9. Figure 4. Back AC power inlet Equipotential stud AC fuses...
  • Page 12 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph Figure 5. Right Side Com port B (USB) SD memory card Reset slot functional button Figure 6. Front Com port A (for patient cable) Figure 7. Left Side Paper tray latch...
  • Page 13 Directions for Use Chapter 1 Introduction Table 1. Keyboard Function A. On/Off “Powering the Electrocardiograph” on page 22. B. Backspace Deletes the character to the left of the cursor. C. Menu “About the Main Menu” on page 10. D. Auto ECG Begins Auto ECGs, normal and stat.
  • Page 14: About The Main Menu

    Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph About the Main Menu The main menu appears when you press the Menu key Figure 8. Main Menu 9:17AM Oct 16 08 Main Menu 1 ECG Settings 2 System Settings 0 Exit...
  • Page 15: Moving Through The Menus

    Directions for Use Chapter 1 Introduction Moving Through the Menus Figure 9. Standard Menu Figure 10. Parent Menu With Submenu 9:17AM Oct 16 08 9:17AM Oct 16 08 Edit Auto Report Format 1 Lead Arrangement 1 Format 3x4 +1R 2 Rhythm Lead 1 2 Interp Settings 3x4 +3R 3 Rhythm Lead 2...
  • Page 16: About The Patient Cable And Leads

    Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph About the Patient Cable and Leads The patient cable processes the patient’s ECG data and transmits it to the electrocardiograph. To make handling convenient, the ten leads are arranged to point toward the appropriate parts of the body. The cable rake, which slides easily, prevents the chest leads from tangling.
  • Page 17: Symbols

    Directions for Use Chapter 1 Introduction Symbols The symbols illustrated on the following pages may appear on the electrocardiograph, on the packaging, on the shipping container, or in this manual. Documentation Symbols WARNING Indicates conditions or practices that could lead to illness, injury, or death.
  • Page 18 Equipment (WEEE). If this product is patient’s heart. Only surface contact with the contaminated, this directive does not apply. patient’s skin is intended.) For more specific disposal information, see www.welchallyn.com/weee, or contact Welch Allyn Customer Service at +44 207 365 6780. Com port B (USB)
  • Page 19: Using The Electrocardiograph Safely

    Directions for Use Chapter 1 Introduction Using the Electrocardiograph Safely Before using or servicing the electrocardiograph, you must read and understand the following safety-related information. General Warnings The following warning statements apply to electrocardiograph use in general. Warning statements that apply specifically to particular procedures, such as connecting the patient cable or performing an ECG test, appear in the corresponding sections of the manual.
  • Page 20 Chapter 1 Introduction Welch Allyn CP 100 Electrocardiograph Warnings Related to Using the Electrocardiograph WARNING This device captures and presents data reflecting a patient’s physiological condition. When reviewed by a trained physician or clinician, this data can be useful in determining a diagnosis. However, the data should not be used as a sole means for determining a patient’s diagnosis.
  • Page 21: General Cautions

    Caution Use only parts and accessories supplied with the device and available through Welch Allyn. The use of accessories other than those specified may result in degraded performance of this device. Caution Portable and mobile RF communications equipment can affect the performance of the electrocardiograph.
  • Page 22: Getting Help

    Ready screen or Lead Status screen for a list of topics available to print. • Review the other information that came with the electrocardiograph. For list, see “Product information” on page 6. • Contact Welch Allyn. For phone numbers, see page...
  • Page 23: Setting Up The Electrocardiograph

    Setting Up the Electrocardiograph Inspecting the Electrocardiograph ........20 Connecting the Patient Cable .
  • Page 24: Inspecting The Electrocardiograph

    Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 100 Electrocardiograph Inspecting the Electrocardiograph Look for obvious signs of shipping damage. If you find any damage, contact Technical Support. For phone numbers, see page 2. Verify that you have received all appropriate options and accessories. See “Options”...
  • Page 25: Loading The Thermal Chart Paper

    Directions for Use Chapter 2 Setting Up the Electrocardiograph Loading the Thermal Chart Paper Squeeze the latch. Pull the paper door to the left. See Figure If any paper remains in the tray, remove it. 2. Remove the outer packaging, including the cardboard bottom, from a new pack of paper.
  • Page 26: Powering The Electrocardiograph

    Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 100 Electrocardiograph Powering the Electrocardiograph The electrocardiograph can run on AC or battery power. WARNING To ensure that electrical safety is maintained when using AC power, the device must be plugged into a hospital-grade outlet.
  • Page 27: Verifying Proper Operation

    Directions for Use Chapter 2 Setting Up the Electrocardiograph To Turn the Electrocardiograph On Press To Turn the Electrocardiograph Off Press and hold. Note If Power-Save is enabled, the electrocardiograph turns off automatically after several idle minutes. To learn how to enable or disable Power-Save, see “Reviewing the Device Configuration Settings”...
  • Page 28: Connecting An External Usb Printer (Optional)

    Chapter 2 Setting Up the Electrocardiograph Welch Allyn CP 100 Electrocardiograph Connecting an External USB Printer (Optional) If desired, you can connect an external printer. It must be a USB inkjet or laser printer that supports PCL (printer control language), such as the following:...
  • Page 29: Reviewing The System Settings

    Reviewing the System Settings “System Settings” Menu Tree ........26 Reviewing the Device Configuration Settings .
  • Page 30: System Settings" Menu Tree

    Chapter 3 Reviewing the System Settings Welch Allyn CP 100 Electrocardiograph “System Settings” Menu Tree System Settings Device Device Configuration Info Set Date/Time About Language Print Settings Date Format Manage Settings Time Format Enable Options Weight Unit Upgrade Software Height Unit...
  • Page 31: Reviewing The Device Configuration Settings

    Directions for Use Chapter 3 Reviewing the System Settings Reviewing the Device Configuration Settings Press the Menu key 2. Choose System Settings > Device Configuration. The following screen appears. Figure 18. “Device Configuration” Screen 9:17AM Oct 16 08 Device Configuration 1 Set Date/Time 2 Language 3 Date Format...
  • Page 32 Chapter 3 Reviewing the System Settings Welch Allyn CP 100 Electrocardiograph Setting (Continued) Description (Continued) Printer Select Internal or External. When “External” is selected — and when an external printer is connected and turned on — reports will print to that printer. For details, see “Connecting...
  • Page 33: Reviewing The Device Information

    Directions for Use Chapter 3 Reviewing the System Settings Reviewing the Device Information Press the Menu key 2. Choose System Settings > Device Info. The following screen appears. Figure 19. “Device Info” Screen 9:17AM Oct 16 08 Device Info 1 About 2 Print Settings 3 Manage Settings 4 Enable Options...
  • Page 34: Transferring A Configuration To Another Electrocardiograph

    Export Configuration Files. The files are copied to the storage device. 4. Remove the storage device, and insert it into another CP 100 electrocardiograph. 5. Choose System Settings > Device Info > Manage Settings > Import Files > Import Configuration Files.
  • Page 35: Reviewing The Ecg Settings

    Reviewing the ECG Settings “ECG Settings” Menu Tree ........32 About Auto ECG Reports.
  • Page 36: Ecg Settings" Menu Tree

    Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph “ECG Settings” Menu Tree ECG Settings Edit Auto Report Miscellaneous Auto Report Settings Lead Configuration Baseline Centering Augmented Electrode Labels Lead Timing Lead Set Format Default Gain Setting Artifact Report...
  • Page 37: About Auto Ecg Reports

    Directions for Use Chapter 4 Reviewing the ECG Settings About Auto ECG Reports An Auto ECG is a report of ECG data in a user-defined format. For an example, see Figure 21. To learn how to set up or interpret a report, see the references on page Figure 21.
  • Page 38 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph Item (in Figure 21 on page 33) Description A. Patient data “Reviewing the Patient Data Fields Available” on page 37. B. ECG measurements Standard. C. Interpretation (optional) “Reviewing the Interpretation and Copy Settings for Auto Reports”...
  • Page 39: Reviewing The Format Settings For Auto Reports

    Directions for Use Chapter 4 Reviewing the ECG Settings Reviewing the Format Settings for Auto Reports Press the Menu key 2. Choose ECG Settings > Edit Auto Report > Format. The following screen appears. Figure 22. Auto Report “Format” Screen 9:17AM Oct 16 08 Format 1 Lead Arrangement...
  • Page 40: Reviewing The Interpretation And Copy Settings For Auto Reports

    Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph Reviewing the Interpretation and Copy Settings for Auto Reports Press the Menu key 2. Choose ECG Settings > Edit Auto Report > Interp Settings. The following screen appears. Figure 23. “Interpretation Settings” Screen...
  • Page 41: Reviewing The Patient Data Fields Available

    Directions for Use Chapter 4 Reviewing the ECG Settings Reviewing the Patient Data Fields Available You can determine which fields appear during patient data entry. To Choose the Fields Press the Menu key 2. Choose ECG Settings > Edit Auto Report > Patient Data. The following screen appears.
  • Page 42 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph Field (Continued) Description (Continued) Yes or no. If yes, this field is enabled for entering blood pressure in standard ### / ### format. Comments Yes or no. If yes, this field is enabled for entering comments.
  • Page 43: Reviewing The Auto Report Settings

    Directions for Use Chapter 4 Reviewing the ECG Settings Reviewing the Auto Report Settings Press the Menu key 2. Choose ECG Settings > Auto Report Settings. The following screen appears. Figure 25. “Auto Report Settings” Screen for ECG Settings 9:17AM Oct 16 08 Auto Report Settings 1 Baseline Centering 2 Lead Timing...
  • Page 44: Reviewing The Miscellaneous Ecg Settings

    Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph Reviewing the Miscellaneous ECG Settings Press the Menu key 2. Choose ECG Settings > Miscellaneous. The following screen appears. Figure 26. “Miscellaneous” Screen for ECG Settings 9:17AM Oct 16 08 These three default settings—gain, baseline...
  • Page 45: Turning The Augmented Pediatric Lead Set On And Off

    Directions for Use Chapter 4 Reviewing the ECG Settings Turning the Augmented Pediatric Lead Set On and Off For a description of the augmented pediatric lead set, see page Press the Menu key 2. Choose ECG Settings > Augmented Lead Set > On or Off. When this option is turned on, the system will prompt you to select the lead set —...
  • Page 46 Chapter 4 Reviewing the ECG Settings Welch Allyn CP 100 Electrocardiograph...
  • Page 47: Performing Ecg Tests

    Performing ECG Tests Connecting the Leads to the Patient ....... . 44 Recording an Auto ECG.
  • Page 48: Connecting The Leads To The Patient

    Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Connecting the Leads to the Patient Help the patient get comfortable. Patient preparation is important for a successful ECG. a. Describe the procedure. If desired, press the Help key and print the page entitled “What Is An ECG?”...
  • Page 49 Directions for Use Chapter 5 Performing ECG Tests Figure 27. Electrode Placement Locations — Adult and Standard Pediatric Electrodes Locations Fourth intercostal space at right sternal border. Fourth intercostal space at left sternal border. yellow yellow Midway between V2 and V4. green green Fifth intercostal space at left of midclavicular line.
  • Page 50 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Figure 28. Electrode Placement Locations — Augmented Pediatric To use the augmented pediatric lead set, you must enable it in the settings. See “Turning the Augmented Pediatric Lead Set On and Off”...
  • Page 51 Directions for Use Chapter 5 Performing ECG Tests 4. If the electrocardiograph’s display is blank, press 5. If prompted to select a lead set type, select Standard or Augmented. For a description of the augmented pediatric lead set, see page 6.
  • Page 52: Recording An Auto Ecg

    Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Recording an Auto ECG An Auto ECG is a report typically showing a 10-second acquisition of 12 leads of ECG information combined with patient data, interpretation, and measurements matrix. To learn how to set up the Auto ECG report format, see “About Auto ECG Reports”...
  • Page 53: Recording A Normal Auto Ecg

    Directions for Use Chapter 5 Performing ECG Tests Recording a Normal Auto ECG For a normal Auto ECG, you enter patient data and do other optional tasks before printing, as shown in Figure 31 on page 48. To Record a Normal Auto ECG Press the Auto ECG key quickly.
  • Page 54 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Figure 33. “Auto ECG Acquisition” Screen Doe, Jane Doe, Jane 9:17AM Oct 16 08 9:17AM Oct 16 08 HR: 75 Auto ECG : Acquisition Auto ECG Acquisition Printing Report The electrocardiograph displays messages here...
  • Page 55 Directions for Use Chapter 5 Performing ECG Tests Figure 34. “Auto ECG Post-Print” Screen Doe, Jane 9:17AM Oct 16 08 Auto ECG : Post-Print 1. Exit 2. Print Copy 3. Repeat Test 4. Export Test 5. Select the desired item. Item Effect Exit...
  • Page 56 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Caution The requirements of AAMI EC11, Section 3.2.7 .2, Frequency and Impulse Response, for an impulse triangle waveform may be impacted by up to 5 milliseconds of small amplitude dampened ringing immediately after the impulse when the muscle filter (35 Hz) is turned on or a small amplitude offset when the baseline filter (0.5 Hz) is turned on.
  • Page 57: Recording A Stat Auto Ecg

    Directions for Use Chapter 5 Performing ECG Tests Recording a Stat Auto ECG A stat Auto ECG is an immediate printout. Stat mode bypasses patient data entry, as shown in Figure 31 on page 48. A temporary ID number is assigned to the patient to identify stat tests. To Record a Stat Auto ECG Press and hold the Auto ECG key The electrocardiograph begins acquiring ECG data.
  • Page 58: Recording A Rhythm Ecg

    Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph Recording a Rhythm ECG A Rhythm ECG is a continuous, real-time printout of three leads at a time. Rhythm ECGs are printouts only. They cannot be sent to an SD memory card or removable USB storage device.
  • Page 59: Adjusting The Ecg Waveforms

    Directions for Use Chapter 5 Performing ECG Tests Adjusting the ECG Waveforms To adjust the waveforms after printing an Auto ECG or while printing a rhythm ECG, press an item’s number key as needed until the desired choice appears. Item Effect Change Leads Cycles through the lead groups three at a time.
  • Page 60 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph...
  • Page 61: Maintaining The Electrocardiograph

    Maintaining the Electrocardiograph Inspecting the Equipment ......... 58 Cleaning the Equipment .
  • Page 62: Inspecting The Equipment

    If alcohol or liquids spill into the electrocardiograph while cleaning, arrange to have it checked before using it again. For Welch Allyn phone numbers, see page Monthly, or more often if needed, follow these cleaning instructions: Disconnect the power plug from the AC outlet.
  • Page 63: Testing The Equipment

    Testing the Equipment WARNING Only qualified service personnel should perform leakage current tests. Whenever the electrocardiograph is serviced or problems are suspected, Welch Allyn recommends the following test procedures: Verify continued electrical safety of the device, using IEC 60601-1 or ANSI/AAMI ES1 methods and limits.
  • Page 64: Recharging A Fully Discharged Battery

    Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph Recharging a Fully Discharged Battery If the electrocardiograph does not turn on when unplugged, the battery may be fully discharged. Note Regardless of the battery condition, you can use the electrocardiograph whenever it is plugged in.
  • Page 65: Replacing The Battery

    Directions for Use Chapter 6 Maintaining the Electrocardiograph Replacing the Battery If you have recharged the battery and the electrocardiograph still does not turn on when unplugged, or if the battery loses its charge quickly, replace the battery as follows. (For part number, see “Accessories”...
  • Page 66 Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph 6. Unplug the two battery connectors. See Figure You may need to pull hard. Use pliers gently if necessary. Figure 41. Battery Connectors Connect a new battery by matching connector sizes. Do not force connectors to mismatch.
  • Page 67: Replacing The Battery (Dc) Fuse

    Directions for Use Chapter 6 Maintaining the Electrocardiograph 9. Replace the battery door and screws. Turn the electrocardiograph rightside-up. It turns on automatically and displays a prompt to enter date and time. 10. Enter the date and time. The electrocardiograph is ready to use. 11.
  • Page 68: Replacing The Ac Fuses

    Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph Replacing the AC Fuses If the green LED on the keyboard does not light up when the electrocardiograph is connected to AC power, you may need to replace one or both of the AC fuses, as follows.
  • Page 69: Storing The Equipment

    Directions for Use Chapter 6 Maintaining the Electrocardiograph Storing the Equipment When storing the electrocardiograph, cords, and accessories, observe the environmental storage conditions. See “Specifications” on page 73. Discarding the Equipment Discard the old battery appropriately. • In the USA, call 1-800-SAV-LEAD for instructions on recycling it. •...
  • Page 70 Chapter 6 Maintaining the Electrocardiograph Welch Allyn CP 100 Electrocardiograph...
  • Page 71: Troubleshooting

    Troubleshooting Problem-Solving Suggestions ........68 Limited Warranty .
  • Page 72: Problem-Solving Suggestions

    Chapter 7 Troubleshooting Welch Allyn CP 100 Electrocardiograph Problem-Solving Suggestions This section includes several tables: • Lead quality problems (Table 2 on page 68) • System failure problems (Table 3 on page 69) • System messages (Table 4 on page 70) If you try these suggestions and still have problems, contact Technical Support.
  • Page 73 Directions for Use Chapter 7 Troubleshooting Table 2. Lead Quality Problems (Continued) Condition Causes Actions AC interference (even-peaked, regular • Electrodes that are dirty, corroded, • Check all electrode contacts and lead wires. voltage superimposed on the waveforms). loose, or positioned on a bony area. •...
  • Page 74 Chapter 7 Troubleshooting Welch Allyn CP 100 Electrocardiograph Table 4. System Messages (alphabetical order) System Message Problems Actions “Insufficient battery power to begin. Battery low or bad. • Recharge the battery. See “Recharging a Fully Please connect AC power and try again.”...
  • Page 75: Limited Warranty

    Product or accessory free of charge. If your Product requires repairs covered by this warranty, upon your request Welch Allyn will loan to you, at no cost, a substitute Product for use until your repaired Product is returned.
  • Page 76: Service Policy

    Welch Allyn CP 100 Electrocardiograph Service Policy All repairs on products under warranty must be performed or approved by Welch Allyn. Unauthorized repairs will void the warranty. In addition, whether or not covered under warranty, any product repair shall exclusively be performed by Welch Allyn certified service personnel.
  • Page 77: A - Specifications

    Specifications Item Specification 16.2 in Dimensions (41.1 cm) 6.2 in (15.7 cm) 15.6 in (39.7 cm) Weight 11.6 lb (5.3 kg) Keyboard type Elastomer keypad with complete alphanumeric keys Paper type 8.25 x 11 inches (21 x 28 cm), Z-fold thermal paper, 200 sheets Thermal printer (internal) Computer-controlled dot array, 8 dots/mm USB printer (external)
  • Page 78 Because of the sampling characteristics and the asynchronism between sample rate and signal rate, digital ECG systems such as the CP 100 and CP 200 may produce a noticeable modulating effect from one cycle to the next, particularly in pediatric recordings. This phenomenon is not physiologic.
  • Page 79: B - Emc Guidance And Manufacturer's Declarations

    Declarations Table 5. Electromagnetic Emissions The CP 100 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 100 electrocardiograph should assure that it is used in such an environment. Emissions Test...
  • Page 80 Welch Allyn CP 100 Electrocardiograph Table 6. Electromagnetic Immunity The CP 100 electrocardiograph is intended for use in the electromagnetic environment specified below. The customer or user of the CP 100 electrocardiograph should assure that it is used in such an environment.
  • Page 81 AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CP 100 electrocardiograph is used exceeds the applicable RF compliance level above, the electrocardiograph should be observed to verify normal operation.
  • Page 82 CP 100 Electrocardiograph The CP 100 electrocardiograph is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the CP 100 electrocardiograph can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CP 100 electrocardiograph as recommended below, according to the maximum output power of the communications equipment.
  • Page 83: Glossary

    Dominant waveforms for all 12 leads. Printed on a separate page, if enabled. CardioPerfect workstation. A PC using Welch Allyn CardioPerfect software. Stores ECG and spirometry test data. Can communicate with other electronic patient-information systems, such as billing and medical records.
  • Page 84 Glossary Welch Allyn CP 100 Electrocardiograph Rhythm ECG. A continuous, real-time printout of ECG data. See also Auto ECG. workstation. See CardioPerfect workstation.
  • Page 85: Index

    AC power. See powering the electrocardiograph confirming. See interpretation of ECGs, automatic: confirming acquisition screen contact information, entering your AHA electrode labels (V1, V2, V3, etc.) contact information for Welch Allyn arrow keys, using contrast, adjusting artifact report Control key (Ctrl)
  • Page 86 Index Welch Allyn CP 100 Electrocardiograph editing Internet address "Auto Report" ECG settings interpretation of ECGs, automatic "miscellaneous" ECG settings about report formats confirming – electrodes ordering labeling schemes. See AHA; IEC as printed on reports ordering settings for Auto Reports...
  • Page 87 Directions for Use Index on/off, manual and automatic safety – options sampling rate, digital ordering information SD memory card. See memory card sending tests to memory card sequential lead timing serial number, displaying service policy pacemakers, use with set-up. See hardware set-up; ECG settings; system settings paper.
  • Page 88 Index Welch Allyn CP 100 Electrocardiograph wall poster, ordering replacement warning symbol defined warranty waveforms, ECG. See leads web site weight-unit setting Welch Allyn Technical Support Welch cups, ordering workstation. See CardioPerfect workstation...
  • Page 90 4341 State Street Road, PO Box 220, Skaneateles Falls, NY 13153-0220 USA 1 800 535 6663, + 1 315 685 4560 www.welchallyn.com Reorder Number (multi-language CD): 401150 Mat. Number (manual only): 708794, Ver: F...